Citra Wahana Sekar Buana

Product

Humidity Instruments

Specification

Relative humidity measurement

 

Measurement range
Accuracy at +20 °C (68 °F)

Factory calibration uncertainty (+20 °C / +68 °F) 
Sensor
Typical long-term stability

0 ... 100 % RH
± 1.0 % RH (0 ... 90 % RH)
± 1.7 % RH (90 ... 100 % RH)
±0.6 %RH (0 ... 40 %RH)
±1.0 %RH (40 ... 97 %RH)
Vaisala HUMICAP® 180R
better than 1 %RH / year

Temperature measurement

 

Measurement range
     HMP75
     HMP76
          short time
     HMP77
Accuracy at +20 °C (+68 °F)
Sensor


-20 ... +60 °C (-4 ... +140 °F)
-50 ... +120 °C (-58 ... +248 °F)
-50 ... +180 °C (-58 ... +356 °F)
-70 ... +180 °C (-94 ... +356 °F)
± 0.2 °C (0.36 °F)
Pt100 RTD Class F0.1 IEC 60751

Operating environment

 

Temperature (MI70 measurement indicator)
Humidity


-10 ... +40 °C (+14 ... +104 °F)
non-condensing

Inputs and outputs

 

Power consumption

Analog output
Menu languages

Rechargeable NiMH battery pack with AC power or 4 x AA alkaline batteries, type IEC LR6
0 ... 1 VDC
English, French, Spanish, German, Swedish, Finnish, Russian, Japanese, Chinese

Mechanics

 

Weight
Housing classification

400 g
IP54

Description

The Vaisala HUMICAP® Hand-Held Humidity and Temperature Meter HM70 is designed for demanding humidity measurement in spot-checking applications. It is also ideal for field checking and calibration of Vaisala's fixed humidity instruments.

  • RH measurement range 0...100 % RH
  • Multilingual user interface
  • Shows measurement trends graphically
  • Proven Vaisala HUMICAP® Sensor technology
  • 3 probes, temperature measurement ranges between -70 and +180 °C
  • Multiprobe operation; dewpoint and CO2 probes can also be connected
  • Displays various humidity parameters
  • Sensor preheat and chemical purge options for demanding conditions
  • Data can be logged, and transferred to a PC via MI70 Link software
  • Rugged and reliable - even in extreme conditions
  • NIST traceable (certificate included)

http://www.vaisala.com/