Citra Wahana Sekar Buana

Product

Barometric Pressure

Specification

Barometric pressure measurement

  Measurement range 50 ... 1100 hPa or 500 ... 1100 hPa

  Serial output  
  Accuracy at +20 °C  
      500 ... 1100 hPa  
         Class A ±0.15 hPa
         Class B ±0.20 hPa
       50 ... 1100 hPa ±0.35 hPa
  Long-term stability  
     500 ... 1100 hPa  
                Class A ±0.1 hPa / year
                Class B ±0.1 hPa / year
     50 ... 1100 hPa ±0.2 hPa / year

  Analog output  
  Accuracy at +20 °C ±0.30 hPa
  Sensor Vaisala BAROCAP®

Operating environment

 

Temperature range
Humidity range

-40 ... +60 °C (-40 ... +140 °F)
non-condensing

Inputs and outputs

  Operating voltage 5 ... 28VDC (reverse polarity protected)
  Outputs RS232C, RS232C/TTL (optional), RS485, non-isolated (optional)

Mechanics

  Weight:  
       instrument 110 g
       cable 28 g/m
  Material (housing) PC plastic
  Electronics housing classification IP65 (NEMA 4)

 

For complete specifications, please see the brochure or user guide in "Documents".

Description

The Vaisala BAROCAP® Digital Barometer PTB210 is ideal for outdoor installations and harsh environments. The PTB210 barometers are used in applications such as weather stations, data buoys and ships, airports, and agrology. They are also an excellent solution for monitoring barometric pressure in industrial equipment such as laser interferometers and engine test benches.

  • 500 ... 1100 hPa or 50 .... 1100 hPa pressure ranges with serial output
  • Different scalings between 500 ... 1100 hPa with analog output
  • Electronics housing IP65 protected against sprayed water
  • Accurate and stable measurement
  • NIST traceable (certificate included)

http://www.vaisala.com/