The
PORTATEST is a fully automatic tester for determination of
the dielectric breakdown voltage of insulating liquid. It
has been designed for laboratory use as well as for field
use for service and commissioning.
The PORTATEST is very comfortable and easy in use. The
requested testing standard may be selected via menu. The
corresponding electrode type as well as the electrode
spacing will be displayed for confirmation before starting
the test
After placing the test vessel with the oil sample into the
test chamber, the protection lid needs to be closed. For
maximum operator security the protection lid is equipped
with a security interlock key switching off the HV when
opened during testing. When starting the test, sample
temperature as well as a customer specific sample number may
be entered and will than be integrated to the test report
printout. The testing sequence according to the selected
testing standard is performed fully automatically.
Common test sequences for national and international
standards are stored in the oil tester. In addition, the
operator can define and store up to 10 freely programmable
test sequences. The following test processes are available:
1.] 12 standard test sequences for the respective standards
2.] 10 freely programmable test sequences
3.] Short test program for a fast evaluation of the
condition
(duration of the test: 5min)
4.] Overload test, single breakdown test
5.] Voltage withstand test (1min … 24h) with programmable
rate of rise, hold voltage and hold time
6.] Step test (1 … 10 steps) with progr. step, rate of rise
and hold times.
Every oil sample is either automatically marked with a
serial number or with a number entered by the operator.
After every test, a detailed test report is printed on the
built-in printer. In addition to the standard used for the
test, the print-out contains the date, the time, the sample
number, individual breakdown values, average value and
standard deviation.
The last 20 test reports will be stored in the tester for
later recall or printout. Additionally test results can be
transmitted to a PC using the optionally available Windows
program PrintCom. With PrintCom, test results can be stored
on the PC and test reports may be printed on the connected
office printer. PrintCom stores the results in EXCEL®
format. The operator may configure an own suitable EXCEL®
form.
|